2.1.1 CRT settings and power switch
(1) Power switch ON/OFF
(2) Inten(sity) control:
Used to adjust the brightness
(3) Focus control:
Adjusts the "trace focus"
(4) Graticule illumination:
Adjusts the variable background illumination
(5) Rotation control:
Used to adjust the trace system to the hori-
zontal and vertical lines on the screen grat-
icule
(6) Voltage selector switch:
Allows correct setting of appropriate mains
voltage
(7) Mains socket:
Used to connect the supplied power cord
(cord with protective conductor for non-
heating appliances)
2.1.2 Vertical amplifier section
(8) CH-1 or X-socket:
Used to connect a measuring signal for
channel 1 or for the X-axis in the case of
X-Y mode
(9) CH-2 or Y-socket:
Used to drive channel 2 or the Y-axis in the
case of X-Y mode
(10) CH 1 AC/GND/DC
Switch for type of coupling for the input
signal to the vertical amplifier of channel 1.
In the AC position, a capacitor that "blocks"
the DC component of the measuring signal
is connected between the input of the
amplifier and the socket.
In the GND position, the amplifier input is
connected to the frame earth
In the DC position, the measuring signal is
fed directly to the signal amplifier input.
(11) CH 2 AC/GND/DC:
Switch for type of coupling for channel 2
(12) CH 1 volts/div:
Used to match the level of the measuring
signal to the input of the amplifier of
channel 1
(13) CH 2 VOLTS/DIV:
Used to match the level of the measuring
signal to the input of the amplifier of
channel 2
(14) and (15) Variable
PULL X 5 MAG:
When these knobs are pulled out, the verti-
cal sweep can be steplessly adjusted bet-
ween the latched-in levels of switches (13)
and (14) up to a ratio of 1:5 (not calibra-
ted). When these knobs are not used, make
sure that they are turned to the right as far
as the stop, latched and pushed in.
120
100
240
220
EXT BLANKING
INPUT
S/N
XXX XXXXX XX XXXXXXX
XXXX XXXXX
WARNING
DISCONNECT SUPPLY
BEFORE CHANGING RANGE
LINE VOLTAGE (50/60 Hz)
FUSE
2 A
1 A
AC 90
~
110 V
AC 108
~
132 V
AC 198
~
242 V
AC 216
~
250 V
AC 100 V
AC 120 V
AC 220 V
AC 240 V
CH1
OUTPUT
CONNECTORCONTROL
RS-232C
34 19 33
6 7