14
2.2.2 Cable connections (signal leads)
Three types of signal leads can be connected to
the oscilloscope:
- a single measurement ribbon cable, insulated
conductor
- a coaxial cable
- or a probe.
A single lead may be sufficient for higher signal
levels of low impedance, such as those of TTL
circuits. This type of signal feed, however, is not
used very often in case of low-level signals
because of the unacceptable distortion it intro-
duces. Unacceptable distortion is caused because
the single lead is not screened.
If signal sources having BNC outputs are to be
connected to the oscilloscope, a coaxial cable
should be used for such connections as a rule.
These cables are screened, i.e. the inner conduc-
tor that carries the signal is screened against
spurious external signals by braided copper
wires or braided tinned copper wires. This
"screen" is connected to the frame earth of the
power source or the oscilloscope.
If signals are to be measured on components or
subassemblies of circuits, probes are used for
this purpose. These are available in various
designs. Some probes have a switch that can be
used to change over to 1:10 attenuation (= cali-
bration setting). Roughly 1/10th of the ampli-
tude of the available measuring signal is displayed
in this switch position.
Example:
Indicated values = 5 mV pp, probe set to 1:10
position ==> actual value = 50 mV pp.
Warning!
Do not exceed the maximum input values.
If the source resistance or the line capacitance
are not known in the case of direct connection
between the device under test and the oscillo-
scope, especially at relatively high frequencies,
use a 1:10 low-capacitance probe.
One method of minimising the measuring error
at high frequencies in the case of coaxial cables
is to use a terminating resistor. The impedance
of this resistor which is connected directly to the
oscilloscope must match the impedance of the
signal source or the lead.
i) If the edges of the rectangular-pulse signal
are over or under modulated, the probe must
be matched to the input capacitance of the
oscilloscope. See figure on the next page.
j) Set the V-MODE switch (18) to channel 2 (CH
2) and repeat steps h) and i) for the other
probe.
h) Set any probe that has a selectable attenu-
ation ratio to 1:10, connect it to the channel
1 input (CH-1) and connect the probe tip to
the CAL output (35). A rectangular-pulse
signal, having an amplitude of 0.5 Vpp
divided by 10 (=Volt peak/peak=ss), should
then appear on the screen.
Kommentare zu diesen Handbüchern