Note:
Before switching on the oscilloscope and after
connecting any peripheral device, always check
that the DIP switch is correctly set in order to
avoid unnecessary errors.
2.4.4 Setting up the oscilloscope (to start
plotter)
Note:
Settings must only be made when the equip-
ment, oscilloscope and peripheral devices are
switched off.
C) Data signalling rate
DIP switches No. 3, 4 and 5 are used to specify the baud rate. The minimum value (slow) is 300 bauds,
the maximum value (fast) is 9600 bauds.
DIP switch No.
ONONON
543
Baud rate
300 BAUD
ONONOFF 600 BAUD
ONOFFON 1200 BAUD
ONOFFOFF 2400 BAUD
OFFONON 4800 BAUD
OFFONOFF 9600 BAUD
OFFOFFON 9600 BAUD
OFFOFFOFF 9600 BAUD
DIP switch 2 Data format
ON 1 start-bit + 8 bit + 1 stop-bit
1 start-bit + 8 bit + 2 stop-bit
OFF
D) Data format settings
DIP switch No. 2 can be used to set the following data formats:
E) Setting the connection type, "send" only or "send" and "receive" = two-way communication
DIP switch 1 is used to specify whether communication between the oscilloscope and peripheral
device (plotter or computer etc.) uses a one-way connection or a two-way connection = "send" and
"receive".
DIP switch 1 Connection type
ON One-way (plotter connection)
Both-way (connection to a PC)
OFF